Multiple Electronic Acupunctoscope (MEA), Model WQ-10C2 is an
instrument used both as a detector of acupuncture points and as an
electrical stimulator for acupuncture therapy and anesthesia. The
WQ-10D1 model is a newly modified unit from the older WQ-10 series
acupunctoscope, featuring a CA(constant amplitude)-AM(amplitude
modulation) selecting switch to allow it to discharge six different
combinations of wave patterns. |
| -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - |
| Electronic Acupunctoscope Feature: |
 |
| WQ-10C2 |
Multiplier Kx1 / Frequency f(c/s) 0~100 / Width T(μs) 250 ~ 300
Multiplier Kx10 / Frequency f(c/s) 0~1000 / Width T(μs) 50 ~ 60 |
| I. |
The adjustment of the stimulating wave forms and frequencies. They are controlled by following knobs and switch: |
| |
F1=fixed frequencies
F2=variable frequencies
K=multiplier
There are two positions on switch k, "x1" and "x0.1". The frequency of
output impulse, F=Kfi, can be adjusted within two different ranges,
(0~100Hz and 0~10Hz). There are 3 different forms of combinations by
using f1 and f2: |
| |
| 1. |
When f2 is at 0, by adjusting f1, a continuous wave is obtained. |
| 2. |
When f2 is at 0, by adjusting f2, an intermittent wave is obtained. |
| 3. |
By adjusting f1 and f2, but keep f1 < f2 , a composite wave of alternating frequencies f1 and f2 is obtained |
|
| |
|
| II. |
Acupuncture therapy and anesthesia |
| |
| 1. |
Set
"Intensity" knob to zero and "E-E or APD" switch to "1.2.3", set
"Hi-Lo" switch to "Hi" or "Lo" position as required. Intensity of "Lo"
position is 1/3 of "Hi". |
| 2. |
Plug output leads into 1.2.3 sockets, then fasten their clips to the handles of the needles. |
| 3. |
Set the desirable waveforms and frequencies with knobs f1, f2 and K. |
| 4. |
Turn "on" main switch |
| 5. |
Adjust intensity to appropriate degree |
|
| III. |
Therapy with Electric, Excitatory(E-E) and 'skin sure' face acupuncture point stimulation: |
|
| 1. |
Use
output leads with banana plugs. Connect the bananas plugs with probes
provided with cup shaped electrodes at terminal ends. Fill the cup with
absorbent cotton soaked with salt solution. Fix the cup electrode onto
the acupuncture points or the other selected parts of the skin. Set
"Hi-Lo" switch to "Hi" position. |
| 2. |
In case of weak stimulation follow the same steps described in II. Plug output leads into 1.2.3 sockets. |
| 3. |
For
strong stimulation, the output will be taken from the "E-E or APD"
socket, which is specially designed for this purpose. Set"E-E or APD"
switch to "E-E1+2+3" position and intensity is regulated by 1.2.3
channels together. |
|
| IV. |
Detection of "Meridian Sensitivity" |
|
Subjects
are found to be more sensitive to acupuncture. A particular sensation
traveling along the pathways of meridians is demonstrated when the
"Ching-Hsueh" (at the terminal parts of four limbs) is stimulated with
electric impulse of appropriate intensity. This sensation is termed the
meridian sensation. In most sensitive subjects the meridian sensation
can travel along the pathway of the meridian and sometimes even extends
to the neighboring meridian . The meridian sensitivity varies with
different subjects, while the intensity of the electric excitation
required to produce this particular sensation varies from person to
person. Subjects who show more "meridian sensitivity" always react to
acupuncture therapy and anesthesia with better results than those who
are insensitive to it. |
|
|
| WQ-10D1 |
| Frequency f(c/s) 0~100 / Width T(μs) 250 ~ 400 |
| I. |
The adjustment of the stimulating wave forms and frequencies. They are controlled by following knobs and switch: |
|
F1=fixed frequencies
F2=variable frequencies
CA (constant amplitude)-AM(amplitude modulation) selecting switch.
CA (constant amplitude)-AM(amplitude modulation) selecting switch.
There are six different forms of combinations by using f1 and f2 CA-AM |
|
| 1. |
When f2 is at 0, by adjusting f1, a continuous (CA) or rise-fall(AM) wave is obtained. |
| 2. |
When f1 is at 0, by adjusting f2, an intermittent (CA) or saw tooth (AM) wave is obtained. |
| 3. |
By
adjusting f1 and f2, but keep f1
|
|
|
 |
| II. |
Acupuncture therapy and anesthesia |
|
| 1. |
Set
"Intensity" knob to zero and "E-E or APD" switch to "1.2.3", set
"Hi-Lo" switch to "Hi" or "Lo" position as required. Intensity of "Lo"
position is 1/3 of "Hi". |
| 2. |
Plug output leads into 1.2.3 sockets, then fasten their clips to the handles of the needles. |
| 3. |
Set the desirable waveforms and frequencies with knobs f1, f2 and K. |
| 4. |
Turn "on" main switch |
| 5. |
Adjust intensity to appropriate degree |
|
| III. |
Therapy with Electric, Excitatory(E-E) and skin sure face acupuncture point stimulation: |
|
| 1. |
Use
output leads with banana plugs. Connect the bananas plugs with probes
provided with cup shaped electrodes at terminal ends. Fill the cup with
absorbent cotton soaked with salt solution. Fix the cup electrode onto
the acupuncture points or the other selected parts of the skin. Set
"Hi-Lo" switch to "Hi" position. |
| 2. |
In case of weak stimulation follow the same steps described in II. Plug output leads into 1.2.3 sockets. |
| 3. |
for
strong stimulation, the output will be taken from the "E-E or APD"
socket, which is specially designed for this purpose. Set"E-E or APD"
switch to "E-E1+2+3" position and intensity is regulated by 1.2.3
channels together. |
|
|
|
|
 |
 |
1. Power Supply ON/OFF
2. Knob Sensitivity of Adjustment
3. Fixed Frequency
4. Variable Frequency Switch, Stimulating Output Socket
5. CA / AM Selecting Switch for WQ-10D1, x0.1 / x1 Selecting Switch for WQ-10C2
6. Frequency Indicator Green LCD Light
7. Adjustable Intensity
8. E - E or APD Socket
9. 10.11. Stimulating Output Socket
12.Inject for AC Adapter ( Adaper not included )
13.Electric Excitatory or Acupuncture Points, Detection Selecting Switch |
|
| -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - |
1. Multiple usages:
A. The section used for acupuncture point detection can also be used
for locating the auriculo-acupuncture points, meridian points, and band
of (?) zone of acupuncture points.
B. The section used for stimulation can also be used for acupuncture
therapy, skin surface acupuncture point stimulation, electrical
Ti-Needle, acupuncture anesthesia, therapy by electric excitation, and
measuring the sensitivity of meridian.
Both A and B sections maybe used at the same time.
2. Multiple transmit messages:
The instrument can transmit three waveforms of impulse series. Their
frequency varies from 0-100 C/S, with several adjustment combinations.
By varying the combinations of these differing massage types, a large
number of stimulating conditions may be derived in order to prevent
repetitious overuse, enhancing therapeutical efficiency.
3.
High intensity of output: divided into "High" and "Low" two ranges,
smooth adjustment, high stability and low power consumption.
4.
Detecting acupuncture points with accuracy and without producing pain
sensation. After the location of point, stimulating pulses can also
provide therapy from the detecting probe. . |
| -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - |
 |
 |
|
WQ-10C2 |
WQ-10D1 |
|
|
| 1. |
Output leads with clips ( 3 sets) |
| 2. |
Output leads with banana plugs ( 1 set) |
| 3. |
Ground pole (1 pc.) |
| 4. |
Cup-shaped electrode for electric excitation (1 pcs.) |
| 5. |
Probe
attachment with a cylindrical electrode for detection of "meridian
sensitivity" at one end and a conical head for acupuncture points
detection at the other ( 1 pc.) |
| 6. |
Monitoring loudspeaker (1 pc.) |
|
| -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - |
| Electronic Acupunctoscope Technical Description: |
| Model name |
WQ-10C2 |
WQ-10D1 |
| Power source |
9V battery, type PP3,6F22 or its equivalent |
9V battery, type PP3,6F22 or its equivalent |
| Output |
3 channels can be in succession connected into one channel as an electric excitation0output |
3 channels can be in succession connected into one channel as an electric excitation0output |
| Multiplier |
x1 |
x10 |
----- |
| Frequency ( c/s ) |
0 ~ 100 |
0 ~ 1000 |
0 ~100 |
| Pluse Width ( s ) |
0.25 ~ 0.4 |
0.4 ~ 0.6 |
0.25 ~ 0.4 |
| Width Type |
continuous, intermittent and dense disperse |
continuous, intermittent and dense disperse |
| Dimensions |
145 x 100 x 70 mm |
145 x 100 x 70 mm |
|
| Available Options: |
| Model: |
|
|
 |
| Shipping Weight: 1.40 kg |
 |
| Shipping Dimension: 30.48 X 20.32 X 10.16 cm |
 |
Ground Shipping Cost: $22.00 USD
Above shipping charge is for USA & Canada only. The international shipping price will be subject to the destination. |
 |
| This product was added to our catalog on Sunday 13 November, 2005. |
| |
|
|
|
 |
 |
|